Siemens SITRANS Probe LU Operating Instructions Manual page 43

Ultraschall transmitter
Hide thumbs Also See for SITRANS Probe LU:
Table of Contents

Advertisement

P052 Vessel Dimension A
Dimension A as used for P050 Vessel Shapes 2, 3, 4, 5, or 7, in the chart on page 36.
Values
Related Parameters
Enter one of the following, using the units selected in P005:
height of the vessel bottom if P050 = 2, 3, 4, or 5
length of one end-section of the vessel if P050 = 7
P053 Vessel Dimension L
Dimension L as used in P050 Vessel Shape 7
Values
Related Parameters
Enter the vessel length L (excluding both end sections) if P050 = 7. Use the units selected
in P005.
P054 Level or Head Breakpoints
P054 and 055 can be used to calculate either level and volume in a vessel, or head level
and flow rate in an open channel device. (For more information on Open Channel
Monitoring, please see Open Channel Monitoring (OCM) on page 91).
Level Breakpoints
When the vessel shape is too complex for any of the preconfigured shapes, you can
define the shape as a series of segments. In P054 you assign a level value to each
breakpoint. In P055 you assign a corresponding volume value to each breakpoint.
Primary Index
Secondary Index
Values
Related Parameters
A5E32337695
SITRANS Probe LU (HART) – OPERATING INSTRUCTIONS
Range
Default
- - - -
P050 Vessel Shape
in the chart on page 36.
,
Range
Default
- - - -
P050 Vessel Shape
P054
Breakpoint number
Range
Default
P055 Volume or Flow Breakpoints
0.0000 to 99999 in units (P005)
0.0
Display when P050 = 0
0.0000 to 99999 in units (P005)
0.0
Display when P050 = 0
0.0000 to 99999 in units (P005)
0.000
Page 37