Siemens SITRANS Probe LU Operating Instructions Manual page 103

Ultraschall transmitter
Hide thumbs Also See for SITRANS Probe LU:
Table of Contents

Advertisement

Wiring setups for hazardous area installations
Always check the device nameplate and process device tag to verify the approval rating.
1. Intrinsically Safe wiring
Device nameplate (ATEX/IECEX/INMETRO/RCM)
SITRANS Probe LU
7MLxxxx-xxxxx-xxxx
Encl.: NEMA/TYPE 4X, 6, IP67, IP68
Amb.Temp.: – 40°C to 80°C
Serial No: GYZ / B1034567
Power Rating: 24V
Nom., 30V
Siemens Milltronics Process Instruments, Peterborough
Assembled in Canada with domestic and imported parts
The ATEX certificate can be downloaded from the product
page of our website at: www.siemens.com/probelu. Go to Support > Approvals /
Certificates.
The IECEx certificate listed on the nameplate can be viewed on the IECEx website. Go to:
http://iecex.iec.ch and click on Ex Equipment Certificates of Conformity then enter the
certificate number IECEx SIR 13.0008X.
Device nameplate (FM/CSA)
SITRANS Probe LU
7ML1234-78910-ABC-D
Serial No: GYZ / S1034567
Power Rating: 24V
Nom., 30V
Siemens Milltronics Process Instruments, Peterborough
Assembled in Canada with domestic and imported parts
FM Intrinsically Safe connection drawing number 23650516, and CSA Intrinsically Safe
connection drawing number 23650517 can be downloaded from the product page of our
website at: www.siemens.com/probelu. Go to Support > Installation Drawings > Level
Measurement > Continuous - Ultrasonic.
For power demands see "Loop Voltage versus Loop Resistance" on page 95.
For wiring requirements: follow local regulations.
Approved dust-tight and water-tight conduit seals are required for outdoor Type 4X /
NEMA 4X, Type 6 / NEMA 6, IP67
Refer to "Instructions specific to hazardous area installations (Reference European
ATEX Directive 2014/34/EU, Annex II, 1/0/6)" on page 98.
2. Non-incendive wiring (FM US only)
SITRANS Probe LU
7ML1234-78910-ABC-D
Serial No: GYZ / S1034567
Nom., 30V
Power Rating: 24V
Siemens Milltronics Process Instruments, Peterborough
Assembled in Canada with domestic and imported parts
FM Class 1, Div 2 connection drawing
number 23650583 can be downloaded from the product page of our website at:
www.siemens.com/probelu. Go to Support > Installation Drawings > Level Measurement >
Continuous - Ultrasonic.
For power demands, see "Loop Voltage versus Loop Resistance" on page 95 of
Appendix A.
A5E32337695
SITRANS Probe LU (HART) – OPERATING INSTRUCTIONS
Max., 4-20 mA
HART
Amb.Temp.: – 40°C to 80°C
Encl.: NEMA/TYPE 4X, 6, IP67, IP68
Max., 4-20 mA
,
Encl.:
NEMA/TYPE 4X, 6, IP67, IP68
Amb.Temp.: – 40°C to 80°C
Max., 4-20 mA
U i = 30 V
I i = 120 mA
P i = 0.8 W
C i = 3.6 nF
L i = 0 mH
ATENÇÃO - RISCO POTENCIAL DE CARGA ELETROSTÁTICA - VEJA INSTRUÇÕES
Class I, Div 1, Group A, B, C, D
Class II, Div 1, Group E, F, G
Class III
Temp. Code: T4
CSA per drawing: 23650517
FM per drawing: 23650516
WARNING: Possible static hazard, do not rub or clean on site.
IP68 locations.
HART
KCC-REM-S49
PROBELU
Ex ia IIC T4 Ga
DNV 12.0070 X
II
1 G
0518
Ex ia IIC T4 Ga
SIRA 06ATEX2355X
IECEx SIR 13.0008X
Vmax = 30 V
HART
Imax = 120 mA
Pmax = 0.8 W
Ci = 3.6 nF
Li = 0.5 mH
159134
Class I,Div.2,
Group A,B,C,D
Temp.Code:T5
Page 97

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents