Hach ORBISPHERE G1100 User Manual page 82

Sensor and analyzer
Table of Contents

Advertisement

Commands
The "Command Output Buffer" is formatted as follows:
Name
Output command toggle (OCT)
Output command ID (OCI)
Output command data byte 1 (OCD1)
Output command data byte 2 (OCD2)
Output command data byte 3 (OCD3)
Output command data byte 4 (OCD4)
The "Command Input Buffer" is located just after the measurement data and is formatted as
follows:
Name
Input command toggle (ICT)
Input command status (ICS)
Input command data byte 1 (ICD1)
Input command data byte 2 (ICD2)
Input command data byte 3 (ICD3)
Input command data byte 4 (ICD4)
The following commands are available:
Change product
Activate sensor (valid for EC sensors only)
Change product command - output
Name
Value
OCT
1-2
OCI
1
OCD1
0
OCD2
0-99
OCD3
0-1
OCD4
Change product command - input
Name
Value
ICT
1-2
ICS
0-3
ICD1
ICD2
ICD3
ICD4
Output byte
Output byte
Output byte
Output byte
Output byte
Output byte
Input byte
Input byte
Input byte
Input byte
Input byte
Input byte
Channel number:
0 = Channel 1
Product number
Erase measurement files:
0 = Never erase the measurement files.
1 = Erase measurement file if necessary (e.g. gas unit changes)
Not used
0 = OK
1 = Unknown command ID
2 = Invalid parameter (e.g. invalid channel no or product number)
3 = Execution failure
Not used
Not used
Not used
Not used
Communication Menu
Type
Size
8 bits
8 bits
8 bits
8 bits
8 bits
8 bits
Type
Size
8 bits
8 bits
8 bits
8 bits
8 bits
8 bits
Comment
Comment
Offset
0
1
2
3
4
5
Offset
74
75
76
77
78
79
81