Inputs/Outputs Menu
The following table lists the default configuration. The first two events on the list are pre-set and
only the priority can be changed:
68
The user may define a maximum of 12 customized events
•
for each analog output and change the order of priority of
all events.
Note: This only applies to Tri linear and Linear outputs. It is not
available if the output characteristic is set to None.
Configure the events that should be signaled at the
corresponding current shown in the right column.
Only one event signal at a time can be sent via the current
•
output. As there is a possibility to have several events at
the same time, an order of priority must be set. This order
has been set by default, but it can be modified to suit
particular needs and conditions. Touch the priority number
in the left column and edit it.
The shaded events in the list have preset outputs and only
•
the priority can be changed. The other events can be
customized by the user. Touch a white text box to call up
the rolling list. Select an event from this list and press OK.
Then adjust the priority as required.
Note: When an event occurs, measurement information is
superseded by the event information on the output.
Table 1 Extended event table
Priority
0
Sensor out
1
Interfering gas error
2
Custom Event 1
3
Custom Event 2
4
Custom Event 3
5
Custom Event 4
6
Custom Event 5
7
Custom Event 6
8
Custom Event 7
9
Custom Event 8
10
Custom Event 9
11
Custom Event 10
12
Custom Event 11
13
Custom Event 12
14
Custom Event 13
Event
I [mA]
0.25
0.50
0.75
1.00
1.25
1.50
1.75
2.00
2.25
2.50
2.75
3.00
3.25
3.50
3.75