Pump auxiliary screen
To get to this screen, select:
Instrument
1
2
Figure 6
Pump auxiliary
1 Primary Flow
2 Column Flow Ready
condition
1100 Series Nano Pump User Manual
➜
➜
More Pump
Pump auxiliary screen
The primary flow is the flow before the EMPV.
Depending on the requirements of the analysis it can be set
to:
• Low solvent consumption mode
• Standard mode
• Fast gradients mode
For more information see
"Hints for Choosing the Primary
Flow"
on page 37.
If the column flow cannot reach the desired setpoint, the
start of the analysis can be interrupted. You may specify the
condition as "%" which determines the percentage range
where the pump is allowed to start the analysis, or "Always
ready", which starts the analysis independent of the column
flow.
Pump auxiliary
Getting started
3
4
Low solvent consumption
200 to 500 µl/min.
Default value
500 to 800 µl/min.
Fast gradients
800 to 1300 µl/min.
2
13