Using The Chemstation To Configure Your Pump; Pump Configuration Screen - Agilent Technologies 1100 Series User Manual

Nano pump
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2
Getting started

Using the ChemStation to Configure your Pump

Pump configuration screen

To get to this screen, select:
Instrument
Step 1
Step 2
Figure 5
Pump configuration
1 Filter volume
The filter volume tells the system the delay volume of the filter. Predefined
volumes or custom volumes can be set by the user
2 Fast composition
Setting Fast composition change/reconditioning to On means that the
change/
system is reconditioned under the 'fast' parameter to return to the initial
reconditioning
composition at the end of the analysis and before the post run. The
parameters are Duration and Maximum Pressure.
Reconditioning is performed by purging the system (pump and inj.) During
this process the needle moves into a waste position. See
the Fast Composition Change/Reconditioning"
3 Mixer volume
In the standard configuration, no mixer is installed in the nano pump
12
More Pump
Pump Configuration

Pump configuration screen

"Hints for Using
on page 30
1100 Series Nano Pump User Manual
Step 3
Small - 20 µl
Standard - 100 µl
<any number> - 0 to 500 µl
Fast System Flush:
0.0 to 99999.9 min. in steps
of 0.1 min.
Column Reconditioning:
0.0 to 99999.9 min. in steps
of 0.1 min.