ResMed ELITE S8 User Manual page 18

S8 series ii positive airway pressure devices
Hide thumbs Also See for ELITE S8:
Table of Contents

Advertisement

Available languages

Available languages

Problem/Possible Cause
Displays error message:
xxxx defines an error)
Component failure.
Displays error message:
Air tubing is loose or blocked.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card:
Data Card is not inserted
correctly.
The Data Card is only partly
inserted.
You may have removed the
Data Card before settings
were copied to the
S8 Escape II.
The card is not a ResScan Data
Card.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card:
There is a data error on the
Data Card.
The following message is NOT displayed on the LCD after you try to update the
settings using the Data Card: Settings Success Remove Card
The settings were not
updated.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card:
The identification details on
the Data Card do not match
the identification details on
your device.
14
SYSTEM ERRxxxx Call Service!
Record error number and call service.
Check tube!! Key if done
Check that the air tubing is connected securely to your
mask and the air outlet to the front of the S8 Series II.
When you have checked the air tubing, press the Start/
Stop key to restart the device. If this does not clear the
message, disconnect the power cord and then
reconnect it to restart the device.
Card Error Remove Card
Check that the patient is inserting the Data Card into the
Data Card module with the arrow facing up, while the S8
Escape II is switched on and not delivering therapy
(standby mode).
Ensure that the Data Card is inserted into the Data Card
slot as far as it will go.
Reinsert the Data Card and wait for the Settings
Success Remove Card message to appear on the
LCD.
Only use a ResScan Data Card.
Settings Error Remove Card
Contact your clinician immediately.
Contact your clinician immediately.
Settings Invalid Remove Card
Contact your clinician immediately.
Solution
or Exxxx (where

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

S8 autoset iiS8 elite ii

Table of Contents