725Ex
Calibration Manual
Cal Volts Input
Connect the UUT as shown in Figure 13.
From the Calibration Menu, type the cal step for Cal Volts Input. The PC displays:
Set the Fluke 5520A to output 0.0000 V, then press the space bar. After a short while, the
following calibration constant and prompt are displayed on the PC:
Set the Fluke 5520A to output 10.0000 V, and press the space bar. After a short while,
the following calibration constants and prompt are displayed on the PC:
Press the space bar to return to the Calibration Menu.
30
Figure 13. Volts Input Calibration Connections
Enter 0 Volts
- press space bar to continue
Offset =
-40
Enter 10.00 Volts
- press space bar to continue
diff = (Counts - Offset)
3032676
=
3032636
Volts per count =
0.000003
- press space bar to continue
-
-40
aal04f.eps