725Ex
Calibration Manual
Cal mA Input
Connect the UUT as shown in Figure 15.
From the Calibration Menu, type the cal step for Cal mA Input. The PC displays:
Set the Fluke 5520A to output 0.000 mA, let the reading settle a few seconds, then press
the space bar on the PC. After a short while, the PC displays the following calibration
constant and new prompt:
Set the Fluke 5520A to output 24.000 mA, let the reading settle a few seconds, then press
the space bar on the PC. After a short while, the UUT displays the following calibration
constants and new prompt:
Press the space bar to return to the Calibration Menu.
32
Figure 15. mA Input Calibration Connections
Enter 0 ma
- press space bar to continue
Offset =
-337
Enter 24.00 ma
- press space bar to continue
diff = (Counts - Offset)
8106924
=
8106587
mA per count =
0.000003
press space bar to continue
-
-337
aal06f.eps