Cal RTD Low Range
Connect the UUT as shown in Figure 21.
From the Calibration Menu, enter the cal step for Cal RTD Low Range. The PC displays:
Apply 15 ohms to 4 wire jacks
press space bar to continue
Set the Fluke 5520A to output 15.00 Ohms, 2-wire output with 2-wire comp off, then
press the space bar on the PC. After a short while, the PC displays:
Apply 350 ohms to 4 wire jacks
press space bar to continue
Set the Fluke 5520A to output 350.00 Ohms then press the space bar on the PC. After a
short while, the Calibrator will return to the Calibration Menu.
Figure 21. RTD Low Range Calibration Connections
Multifunction Process Calibrator
Calibration Adjustment Procedures
aal09f.eps
39