Lower Display Ma Measurement Tests - Fluke 725Ex Calibration Manual

Multifunction process calibrator
Hide thumbs Also See for 725Ex:
Table of Contents

Advertisement

Lower Display mA Measurement Tests

1. Press RESET on the 5520A.
2. Press l on the UUT until MEASURE and mA appear on the lower display.
3. Make the connections shown in Figure 5.
4. Set up the 5520A to output each of the voltages shown in Table 8 and verify that the
UUT readings are within the limits shown.
5. Press STBY on the 5520A.
5520A Outputs
0.000 mA
12.000 mA
24.000 mA
Figure 5. Lower Display mA Test Connections
Table 8. Lower Display mA Readings
Multifunction Process Calibrator
UUT Readings
-0.002 mA to +0.002 mA
11.995 mA to 12.005 mA
23.993 mA to 24.007 mA
Performance Tests
aal06f.eps
17