Lower Display Ohms Source Tests - Fluke 725Ex Calibration Manual

Multifunction process calibrator
Hide thumbs Also See for 725Ex:
Table of Contents

Advertisement

Lower Display Ohms Source Tests

1. Press F on the UUT until Ω appears on the lower display. If necessary, press M
until SOURCE appears on the lower display.
2. Set the HP 3458A to measure 4-wire resistance.
3. Make the connections shown in Figure 12.
4. Use the arrow keys on the UUT to set the UUT output to the resistance values in
Table 17 and verify that the HP 3458A readings are within the limits shown.
Figure 12. Lower Display Ohms Source Test Connections
UUT Settings
15.0 Ω
360.0 Ω
1500 Ω
3200 Ω
Table 17. Lower Display Ohms Source Readings
Multifunction Process Calibrator
HP 3458A Readings
14.9 Ω to 15.1 Ω
359.9 Ω to 360.1 Ω
1499.5 Ω to 1500.5 Ω
3199.0 Ω to 3201.0 Ω
Performance Tests
aal13f.eps
27