Specifications; Dc Voltage Measurement; Dc Voltage Source; Millivolt Measurement And Source - Fluke 725Ex Calibration Manual

Multifunction process calibrator
Hide thumbs Also See for 725Ex:
Table of Contents

Advertisement

Specifications

All specifications apply from +18 °C to +28 °C unless stated otherwise. All specifications
assume a five-minute warmup period.

DC Voltage Measurement

Range
30 V (upper display)
10 V (lower display)
90 mV
Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ±0.005 % of range per °C

DC Voltage Source

Range
100 mV
10 V
Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ±0.005 % of range per °C
Maximum load: 1 mA

Millivolt Measurement and Source*

Range
-10 mV to 75 mV
Maximum input voltage: 30 V
Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ±0.005 % of range per °C
*Select this function by pressing T. The signal is available at the thermocouple miniplug connector.

DC mA Measurement and Source

Range
24 mA
Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ±0.005 % of range per °C
Drive capability: 250 Ω at 20 mA
Resolution
0.001 V
0.001 V
0.01 mV
Resolution
0.01 mV
0.001 V
Resolution
0.01 mV
Resolution
0.001 mA
Multifunction Process Calibrator
Accuracy,
(% of Reading + Counts)
0.02 % + 2
0.02 % + 2
0.02 % + 2
Accuracy,
(% of Reading + Counts)
0.02 % + 2
0.02 % + 2
Accuracy
±(0.025 % + 1 count)
Accuracy,
(% of Reading + Counts)
0.02 % + 2
Specifications