Lower Display mV Source Tests

1. Press M on the UUT until SOURCE appears on the lower display; then press V
2. Set the HP 3458A to measure dc voltage in the 200 mV range.
3. Connect the UUT to the HP 3458A as shown in Figure 7.
4. Use the arrow keys on the UUT to set the UUT output to the current values in Table
5. Press Mon the UUT to disable the sourcing function.

Lower Display Voltage Source Tests

1. Press M on the UUT until SOURCE appears on the lower display; then press V
2. Set the HP 3458A to measure dc voltage in the 20 V range.
3. Connect the UUT to the HP 3458A as shown in Figure 7.
4. Use the arrow keys on the UUT to set the UUT outputs to the currents in Table 16
26
until mV appears on the lower display.
15 and verify that the HP 3458A readings are within the limits shown.
Table 15. Lower Display mV Source Readings
UUT Outputs
0.00 mV
50.00 mV
100.00 mV
until V appears on the lower display.
and verify that the HP 3458A readings are within the limits shown. A lower voltage
range on the HP 3458A may be used to verify the 0 V range.
Table 16. Lower Display Voltage Source Readings
UUT Outputs
0.000 V
5.000 V
10.000 V
HP 3458A Readings
-0.020 mV to +0.020 mV
49.970 mV to 50.030 mV
99.960 mV to 100.040 mV
HP 3458A Readings
-0.002 V to +0.002 V
4.9970 V to 5.0030 V
9.9960 V to 10.0040 V