Cal Thermocouples

Connect the UUT as shown in Figure 19.
From the Calibration Menu, enter the cal step for Cal Thermocouples. The PC displays:
Set the Fluke 5520A to output 0.00 degrees C for a type-J thermocouple using the ITS-90
standard, then press the space bar on the PC. After a short while, the PC displays the
following calibration constant and new prompt:
Press the space bar to return to the Calibration Menu.
36
Figure 19. Thermocouples Calibration Connections
Connect accurate T/C source : TYPE J thermocouple
Simulate 0.0 degrees C
- press space bar to continue
CJC Temp Read = 27.359071
Type
2
CJC mV =
1.400393
mV Read = -1.260579
CJC = 0.139814,
(cjc - type_j) =
- press space bar to continue
2.736758
aal11f.eps