SECTION 6—OPERATING INSTRUCTIONS
Your oxygen concentrator will perform best when operated
under the conditions outlined in the Typical Product
Parameters on page 13. Usage in environments other than
those described may result in the need for increased
equipment maintenance. The air intake of the unit should be
located in a well ventilated area to avoid airborne pollutants
and/or fumes.

Set Up

1. Plug in power cord to an electrical outlet.
2. Connect Humidifier (If So Prescribed).
DO NOT fill humidifier bottle with hot water. Allow boiled
water to cool to room temperature before filling.
DO NOT overfill humidifier.
DO NOT reverse the oxygen input and output connections.
Water from the humidifier bottle will travel through the
cannula back to the patient.
When using tubing connections longer than 2.1 m (7 feet) in
length, position the humidifier as close to the patient as
possible to allow for maximum humidification output.
NOTE: For this procedure, refer to FIGURE 6.1, FIGURE 6.2 on
page 19, FIGURE 6.3 on page 21.
1. Remove cap from bottle.
2. Fill humidifier with boiled tap water or bottled water to
the level indicated by the manufacturer. Boil tap water for
approximately 10 minutes and cool to room temperature
prior to use. Replace the humidifier cap and securely
tighten.
Perfecto
™Series
2
WARNING
18
Part No 1163145