SECTION 5—TYPICAL PRODUCT PARAMETERS
Altitude:
Oxygen Output
Concentration
Levels:
Maximum Outlet
Pressure:
Low Flow Alarm:
Potential
Obstruction Alert
Power
Consumption:
Pressure Relief
Mechanism
Operational at:
Change in
maximum
recommended
flow when back
pressure of 7kPa is
applied:
Filters:
Safety System:
Width:
Height:
Depth:
Perfecto
™Series
2
Up to 6,000 ft (1828 meters) above sea level without degradation
of concentration levels. Not recommended for use above 6000 ft
(1828 meters).
Atmospheric pressure range: 101.33kPa - 81.22kPa
87% to 95.6% @ 0.5 to 5 L/min
93% ± 3% @ 2 L/min
NOTE: Stated concentration levels achieved after initial warm-up
period (approximately 30 minutes).
34.5 kPa ± 3.45 kPa (5 psi ± 0.5 psi)
0.5 to 5 L/min (maximum). For flowrates less than 1 L/min, we
recommend the use of the Invacare Pediatric Flowmeter
Accessory (IRCPF16AW).
0 L/min to 0.5 L/min
The concentrator detects a condition that may indicate a
potential obstruction of the output oxygen.
Rapid audible beeping alert (this alert is deactivated when
accessories are connected). May be associated with flow setting
of 0.5 L/min or less.
Perfecto
AW - 300 W @ 5 L/min, 280 W @
2
Perfecto
VAW - 320 W @ 5 L/min
2
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)
0.7 L/min
Cabinet, Outlet HEPA and Compressor Inlet
Current overload or line surge shutdown. High temperature
compressor shutdown. High Pressure Alarm w/compressor
shutdown. Low Pressure Alarm w/compressor shutdown.
Battery Free Power Loss Alarm. SensO2 Oxygen System
Possible Obstruction Alert.
38.1 cm ± 1 cm (15 in ± 3/8 in)
58.4 cm ± 1 cm (23 in ± 3/8 in)
30.5 cm ± 1 cm (12 in ± 3/8 in)
14
3 L/min
Part No 1163145