Cleaning and Disinfection Between Patients
ONLY qualified personnel should perform cleaning and
disinfection of the concentrator and accessories between
patients.
NOTE: Follow these instructions to eliminate possible pathogen
exchange between patients due to contamination of components or
accessories. Preventive Maintenance should also be performed at
this time if necessary.
1. Dispose of and replace all patient side accessories not
suitable for multiple patient use, including but not limited
to:
• Nasal Cannula and Tubing
• Mask
• Humidifier
2. Perform Routine Maintenance on page 25 and Preventive
Maintenance on page 28, as applicable.
3. Check concentrator for possible external damage or signs
that may require service or repair.
4. Ensure concentrator functions properly and all alarms are
in working order.
5. Before repackaging and distribution to new patient,
ensure packaging contents contain the concentrator,
power cord, air inlet scoop, assembly instructions,
humidifier, cannula, labels and user manual.
Part No 1163145
SECTION 7—MAINTENANCE