SECTION 7—MAINTENANCE
SECTION 7—MAINTENANCE
WARNING
The Invacare concentrators are specifically designed to
minimize routine preventive maintenance. Only qualified
personnel should perform preventive maintenance on the
concentrator.
Unplug the concentrator when cleaning. To avoid electrical
shock, DO NOT remove cabinet.
NOTE: At a minimum, preventive maintenance MUST be
performed according to the maintenance record guidelines. In places
with high dust or soot levels, maintenance may need to be performed
more often. Refer to Preventive Maintenance Record on page 28.

Routine Maintenance

Cleaning the Cabinet Filter
CAUTION
DO NOT operate the concentrator without the filter installed.
NOTE: For this procedure, refer to FIGURE 7.1 on page 26.
NOTE: There is one cabinet filter located on the back of the cabinet.
1. Remove the filter and clean as needed.
NOTE: Environmental conditions that may require more frequent
inspection and cleaning of the filter include, but are not limited to:
high dust, air pollutants, etc.
2. Clean the cabinet filter with a vacuum cleaner or wash in
warm soapy water and rinse thoroughly.
3. Dry the filter thoroughly before reinstallation.
Part No 1163145