3.4 Configuration

Lid/Cover Removal

To switch on your unit and access the USB programming connector on your
meter, the enclosure lid/cover must be removed. Four screws must be removed.
Be careful to not lose or discard these screws. These screws play a vital part in
providing the water tight seal on your unit.
The same four screws removed in this procedure must be
re-installed into your meter housing or you will not have the
proper water tight seal. Failure to install these screws correctly
may result in damage to your unit when the meter is exposed
to wet conditions.
When reinstalling enclosure screws they must be tightened to
9 in-lb ±10%.
REMOVE THESE FOUR SCREWS
THE METER FRONT COVER
IN ORDER TO REMOVE
Figure 3-6. Lid/Cover Removal