Download Print this page
Omega DPG1001AD Instruction Sheet

Omega DPG1001AD Instruction Sheet

Digital pressure gauges

Advertisement

Quick Links

Digital Pressure Gauges
DPG1000AD, DPG1001AD, DPG1000ADBL,
DPG1100AD, DPG1100ADBL
INSTRUCTION
M3361/0606
SHEET
Shop online at: omega.com e-mail: info@omega.com
For latest product manuals: omegamanual.info
DPG1000AD
DPG1001B
DPG1000ADBL Low Voltage AC/DC Powered w. Backlighting
DPG1100AD
DPG1100ADBL Low Voltage AC/DC Powered w. Backlighting, NEMA 4X
Ranges and Resolution
Resolution is fixed as indicated in table
Ranges in Bold are standard. Contact Omega Engineering to order other ranges listed below.
abs:
Absolute reference (normally reads atmospheric pressure, reads zero at full vacuum)
vac:
Vacuum gauge, minus sign not used unless specified
–30.0 inHg/15.0 psig
120.0 inHg
–30.0 inHg/100.0 psig
199.9 inHg abs
–30.0 inHg/199.9 psig
199.9 inHg
3.00 psig
50.0 oz/in2
5.00 psig
80.0 oz/in2
15.00 psi abs
240 oz/in2 abs
15.00 psig vac
240 oz/in2 vac
±15.00 psig
±240 oz/in2
15.00 psig
240 oz/in2
30.0 psi abs
85.0 inH
30.0 psig
140.0 inH
60.0 psig
400 inH
100.0 psi abs
400 inH
100.0 psig
±400 inH
199.9 psig
400 inH
300 psig
850 inH
500 psig
7.00 ftH
1000 psig
12.00 ftH
1999 psig
35.0 ftH
3000 psig
70.0 ftH
5000 psig
140.0 ftH
6.00 inHg
230 ftH
10.00 inHg
480 ftH
30.0 inHg abs
150.0 mmHg
30.0 inHg vac
260 mmHg
±30.0 inHg
760 mmHg abs
30.0 inHg
760 mmHg vac
60.0 inHg abs
760 mmHg
60.0 inHg
1600 mmHg abs
Accuracy
(linearity, hysteresis, repeatability)
±0.25% of full scale ±1 least significant digit
DPG1001AD: ±0.10% of full scale ±1 least
significant digit
Display
(update rate, type, size)
3 readings per second nominal display update
Ranges up to 1999: 3
1
/
digit LCD,
2
Ranges 2000 and up: 4 digit LCD, 0.4" H digits,
alphanumeric display of units
BBL: Red LED display backlighting
Controls & Functions
Ranges up to 1999 : Front pushbutton turns
gauge on/off.
Ranges 2000 and up: Front pushbutton turns
gauge on/off. Pressing and holding front button
zeros gauge. The zero feature is not used on
absolute reference gauges.
Low Voltage AC/DC Powered
Low Voltage AC/DC Powered, High Accuracy
Low Voltage AC/DC Powered, NEMA 4X
1600 mmHg
760 torr abs
1600 torr abs
2100 mmH
O
2
3500 mmH
O
2
199.9 cmH
O
2
350 cmH
O
2
1000 cmH
O
2
2100 cmH
O
2
O
199.9 mbar
2
O
350 mbar
2
O abs
1000 mbar abs
2
O vac
1000 mbar vac
2
O
±1000 mbar
2
O
1000 mbar
2
O
2000 mbar abs
2
O
2000 mbar
2
O
4000 mbar
2
O
1.000 bar abs
2
O
1.000 bar vac
2
O
±1.000 bar
2
O
1.000 bar
2
O
1.999 bar abs
2
1.999 bar
4.00 bar
7.00 bar abs
7.00 bar
14.00 bar
19.99 bar
Calibration
Ranges up to 1999: Front potentiometers, non-
interactive zero & span, ±10% range
Ranges 2000 and up: Internal pushbuttons, non-
interactive zero, midpoint, & span, ±10% range.
Calibration information stored in non-volatile
memory.
1
/
" digit height
2
Connection Size
1
/
" NPT male
4
Sensor Material
All wetted parts are 316 stainless steel
Power
AD: Approximately 5 mA
ADBL: Approximately 80 mA
Operates on 8 to 24 VAC 50/60 Hz or 9 to 32
VDC
Optional wall mount power supply is available to
operate on 115VAC
DPG1000AD,
DPG1001AD,
DPG1000ADBL
35.0 bar
1.000 kg/cm
2
70.0 bar
1.000 kg/cm
2
140.0 bar
±1.000 kg/cm
2
199.9 bar
1.000 kg/cm
2
350 bar
1.999 kg/cm
2
19.99 kPa
1.999 kg/cm
2
35.0 kPa
4.00 kg/cm
2
100.0 kPa abs
7.00 kg/cm
2
100.0 kPa vac
7.00 kg/cm
2
±100.0 kPa
14.00 kg/cm
2
100.0 kPa
19.99 kg/cm
2
199.9 kPa abs
35.0 kg/cm
2
199.9 kPa
70.0 kg/cm
2
400 kPa
140.0 kg/cm
2
700 kPa abs
199.9 kg/cm
2
700 kPa
350 kg/cm
2
1500 kPa
1.000 atm abs
1999 kPa
±1.000 atm
3500 kPa
1.000 atm
5000 kPa
4.00 atm
3.50 MPa
7.00 atm
7.00 MPa
14.00 atm
14.00 MPa
19.99 atm
19.99 MPa
35.0 atm
35.0 MPa
70.0 atm
1000 g/cm
2
abs
135.0 atm
1000 g/cm
2
199.9 atm
2100 g/cm
2
abs
340 atm
2100 g/cm
2
DPG1100AD,
DPG1000AD,
DPG1100ADBL
DPG1000ADBL
Ranges 2000
and up
Dimensions
(see drawing on next page)
DPG1000, DPG1001
3.5" W x 3.0" H x 2.0" D NEMA 4X housing
Add approx. 0.75" to height for pressure fitting
DPG1100
3.38" W x 2.88" H x 1.65" D housing
Add approx. 0.75" to height for pressure fitting
Weight
(approximate)
Gauge:
9 ounces
Shipping weight: 1 pound
abs
vac
Housing Material & Color
DPG1000, DPG1001
Extruded aluminum case, light gray epoxy
abs
powder coated, black ABS/polycarbonate bezel,
front and rear gaskets, polycarbonate label
DPG1100
abs
Light gray ABS/polycarbonate NEMA 4X case,
rear gasket, polycarbonate label
Overpressure
2 times sensor range
3000 psig range & metric equivalents: 5000 psig
5000 psig range & metric equivalents: 7500 psig
Ranges 2000 and up: 112.5% out-of-range
display: i – – – or i –.–.–.– depending on model
Burst Pressure
4 times sensor pressure rating, or 10,000 psi,
whichever is less
Storage Temperature
–40 to 203°F (–40 to 95°C)
Operating Temperature
–4 to 185°F (–20 to 85°C)
Compensated Temperature
32 to 158°F (0 to 70°C)
Temperature Stability
±0.5% span typical, ±1% span max. 0 to 82°C
Optional Power Supply Kit
DPG1000-PS
The kit includes a UL and CSA listed 115 VAC
(50/60 Hz) wall-mount power supply with U.S.-
style 2-prong plug. Output is 12 VDC at 200
DPG1100AD,
DPG1100ADBL
Ranges 2000
and up
mA.
The
two-
conductor wire is
approximately 6 feet
long with stripped
wire ends. Use a
pair of pliers to
connect gauge and
power supply wires
using the included
moisture-resistant
connector.

Advertisement

loading

Summary of Contents for Omega DPG1001AD

  • Page 1 3.38" W x 2.88" H x 1.65" D housing Resolution is fixed as indicated in table Add approx. 0.75" to height for pressure fitting Ranges in Bold are standard. Contact Omega Engineering to order other ranges listed below. Weight (approximate)
  • Page 2 0.75" 0.75" PSIG PSIG Zero Span Zero Span OMEGA Engineering Inc. / Made in USA OMEGA Engineering Inc./Made in USA Turn at hex Turn at hex fitting only! fitting only! 0.75" 0.75" "NPT "NPT 3.38 "...
  • Page 3 All gauges are factory calibrated on NIST traceable calibration equipment. No calibration is required before placing the gauge into service. Gauges can be returned to Omega Engineering for factory certified recalibration. NIST traceability is available. Gauges can also be recalibrated...
  • Page 4 European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.

This manual is also suitable for:

Dpg1000adDpg1000adblDpg1100adDpg1100adbl