Omega DPG409 SERIES User Manual page 29

High accuracy digital pressure gauge
Table of Contents

Advertisement

You must set the wire jumpers located on the back of your
meter to match the analog output type you selected here in
the setup process.
TB2
BATTERY
S1
Scaling
Here you can scale the analog output to correspond to the process reading value.
4-20 mA Example:
This table shows analog output values you should expect if you set 4mA = 0 psig
and 20mA = 1000 psig.
Process Reading
0 psi
250 psi
500 psi
750 psi
1000 psi
NOTE:
BATTERY
TB1
Figure 3-18. Analog Output Options
Analog Output Value
4.00 mA
8.00 mA
12.00 mA
16.00 mA
20.00 mA
Setup & Configuration
0-5 Vdc/
4-20 mA SETTING
0-10Vdc SETTING
A
A
B
B
C
C
3
3-12

Advertisement

Table of Contents
loading

Table of Contents