Endress+Hauser Proline Promass series Brief Operating Instructions

Part 1 of 2 coriolis sensor

Advertisement

Quick Links

KA01212D/06/EN/03.17
71360078
Products
Brief Operating Instructions
Proline Promass
Part 1 of 2
Coriolis sensor
These Instructions are Brief Operating Instructions; they are
not a substitute for the Operating Instructions pertaining to
the device.
These Brief Operating Instructions contain all of the
information on the sensor. Please also follow the Brief
Operating Instructions for the transmitter during
commissioning →  3.
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Promass series

  • Page 1 Products Solutions Services KA01212D/06/EN/03.17 71360078 Brief Operating Instructions Proline Promass Part 1 of 2 Coriolis sensor These Instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. These Brief Operating Instructions contain all of the information on the sensor.
  • Page 2: Serial Number

    Proline Promass Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 3 These Brief Operating Instructions are the Sensor Brief Operating Instructions. The "Transmitter Brief Operating Instructions" are available via: • Internet: www.endress.com/deviceviewer • Smart phone/tablet: Endress+Hauser Operations App Detailed information about the device can be found in the Operating Instructions and the other documentation: • Internet: www.endress.com/deviceviewer...
  • Page 4 Disposing of the measuring device ............31 Endress+Hauser...
  • Page 5: Document Information

    Result of a step Visual inspection 1.1.3 Electrical symbols Symbol Meaning Symbol Meaning Direct current Alternating current Direct current and alternating current Ground connection A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system. Endress+Hauser...
  • Page 6 Allen key Open-ended wrench 1.1.6 Symbols in graphics Symbol Meaning Symbol Meaning 1, 2, 3,... Item numbers … Series of steps A, B, C, ... Views A-A, B-B, C-C, ... Sections Hazardous area Safe area (non-hazardous area) Flow direction Endress+Hauser...
  • Page 7: Basic Safety Instructions

    Danger of breakage due to corrosive or abrasive fluids! ‣ Verify the compatibility of the process fluid with the sensor material. ‣ Ensure the resistance of all fluid-wetted materials in the process. ‣ Keep within the specified pressure and temperature range. Endress+Hauser...
  • Page 8: Workplace Safety

    Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 9: It Security

    Incoming acceptance and product identification Incoming acceptance Are the order codes on the delivery note (1) and the product sticker A0028673 (2) identical? A0029314 A0029315 Are the goods undamaged? A0028673 A0029316 Endress+Hauser...
  • Page 10: Product Identification

    A0029318 • If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center. • Depending on the device version, the CD-ROM might not be part of the delivery! The Technical Documentation is available via the Internet or via the Endress+Hauser Operations App.
  • Page 11: Storage And Transport

    ‣ Protect from direct sunlight to avoid unacceptably high surface temperatures. ‣ Store in a dry and dust-free place. ‣ Do not store outdoors. Transporting the product Transport the measuring device to the measuring point in the original packaging. Endress+Hauser...
  • Page 12 The device must always be secured at two lifting lugs at least. 4.2.3 Transporting with a fork lift If transporting in wood crates, the floor structure enables the crates to be lifted lengthwise or at both sides using a forklift. Endress+Hauser...
  • Page 13: Installation

    However, the following installation suggestion allows for installation in an open vertical pipeline. Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter prevent the sensor running empty while measurement is in progress. Endress+Hauser...
  • Page 14 [in] [mm] [in] ¹⁄₂₄ 0,03 ¹⁄₁₂ 0,06 ¹⁄₈ 0,12 ³⁄₈ 0,24 ½ 0,40 15 FB ½ FB 0,60 0,55 25 FB 1 FB 0,95 1½ 0,87 40 FB 1½ FB 1,38 1,10 50 FB 2 FB 2,13 1,97 2,60 Endress+Hauser...
  • Page 15 Applications with high process temperatures may increase the ambient temperature. To maintain the maximum ambient temperature for the transmitter, this orientation is recommended. Promass A, E, F, G, O Promass X Promass H, I, P, Q, S Endress+Hauser...
  • Page 16 For detailed information on the ambient temperature range, see the Operating Instructions for the device. If operating outdoors: Avoid direct sunlight, particularly in warm climatic regions. Temperature tables For detailed information on the temperature tables, see the separate document entitled "Safety Instructions" (XA) for the device. Endress+Hauser...
  • Page 17 Ensure that convection takes place on a sufficiently large scale at the transmitter neck. ‣ Ensure that a sufficiently large area of the housing support remains exposed. The uncovered part serves as a radiator and protects the electronics from overheating and excessive cooling. Endress+Hauser...
  • Page 18 Promass P: For the extended temperature range, version with long extension neck, order code for "Measuring tube material", option TD, TG or extension neck for insulation, order code for "Sensor option", option CG Promass I and S: For the version extension neck for insulation order code for "Sensor option", option CG Endress+Hauser...
  • Page 19 Proline Promass Installation 40(104) 60(140) [°C] [°F] A0029981  5 Valid for Promass I, S 40(104) 60(140) [°C] [°F] A0029990  6 Valid for Promass X Endress+Hauser...
  • Page 20 Valid for Promass F Insulation thickness Medium temperature Maximum recommended insulation thickness at an ambient temperature of T = 40 °C (104 °F) 40(104) Maximum recommended insulation thickness at an ambient temperature of T = 60 °C (140 °F) 60(140) Endress+Hauser...
  • Page 21 Minimum distance to insulation maximum Insulation thickness The minimum distance a between the transmitter sensor connection housing and the insulation is 10 mm (0,39 in) 20 mm (0,79 in). This is to ensure that the transmitter sensor connection housing remains completely exposed. Endress+Hauser...
  • Page 22 Make sure that the function and operation of the rupture disk is not impeded through the installation of the device. The position of the rupture disk is indicated on a sticker beside it. The transportation guard must be removed. Endress+Hauser...
  • Page 23 Make sure that the function and operation of the rupture disk is not impeded through the installation of the device. The position of the rupture disk is indicated on a sticker applied over it. If the rupture disk is triggered, the sticker is destroyed. The disk can therefore be visually monitored. Endress+Hauser...
  • Page 24 In the event of a failure of the rupture disk, a discharge device can be screwed onto the internal thread of the rupture disk in order to drain off any escaping medium. Endress+Hauser...
  • Page 25 It is not necessary to provide additional support for the sensor for operational performance purposes. If, however, additional support is required for installation purposes, the following dimensions must be observed. Use mounting clamp with lining between clamp and measuring instrument. A0016588 Endress+Hauser...
  • Page 26 The sensor must be mounted in the correct position to ensure full drainability in the horizontal position. Markings on the sensor show the correct mounting position to optimize drainability. Promass I Endress+Hauser...
  • Page 27 "This side up" label indicates which side is up Slope the device in accordance with the hygiene guidelines. Slope: approx. 2 ° or 35 mm/m (0.42 in/ feet) Line on the underside indicates the lowest point of the eccentric process connection. Transmitter Endress+Hauser...
  • Page 28 • Device supported on an angle bracket mounted on the wall A0019631 Horizontal Device standing on a solid support base A0019632 Post retainer (Promass A) The post retainer mounting kit is used to secure the device to a pipe or post (order code for "Accessories", option PR). Endress+Hauser...
  • Page 29: Mounting The Measuring Device

    Required tools For transmitter • For turning the transmitter housing: Open-ended wrench8 mm • For opening the securing clamps: Allen key3 mm • For turning the transmitter housing: Open-ended wrench8 mm • For opening the securing clamps: Allen key3 mm Endress+Hauser...
  • Page 30 Ensure that the direction of the arrow on the nameplate of the sensor matches the flow direction of the fluid. Install the measuring device or turn the transmitter housing so that the cable entries do not point upwards.  A0029263 Endress+Hauser...
  • Page 31: Post-Installation Check

    Ensure that the measuring device and all cavities are free of fluid residues that are hazardous to health or the environment, e.g. substances that have permeated into crevices or diffused through plastic. Observe the following notes during disposal: ‣ Observe valid federal/national regulations. ‣ Ensure proper separation and reuse of the device components. Endress+Hauser...
  • Page 32 www.addresses.endress.com...