XNX Universal Transmitter
6.2.2 EC Sensor Performance Data, Factory Mutual Verified (see Section 6.3)
Selectable Full
Scale Range (Dis-
Gas
Cartridge P/N
play and 4-20mA
O
Oxygen
XNXXSO1FM
2
H
S
Hydrogen Sulfide
XNXXSH1FM
10.0 to 50.0 ppm
2
XNXXSC1FM 1
CO
Carbon Monoxide
100 to 1000 ppm
FOOTNOTES:
1. XNXXSC1FM accuracy over temperature <±10% of reading 20°C/68°F to 55°C/131°F, <±20% of reading 20°C/68°F to -10°C/14°F, <±30% of reading -10°C/14F to -20°C/-4°F�
Recalibration is recommended if the temperature of the local environment has varied by more than -30°C�
NOTES:
• Performance figures are measured by test units calibrated at 50% of full scale, at ambient conditions of 20°C, 50% RH, with the EC weatherproof cover attached
• IP rating of FM Cartridges is IP63�
• Barometric pressure effects on the O
pressure changes by ±20% the output from the O
ambient pressure due to either weather or altitude� For example, a 10KPa instantaneous positive pressure step change may cause an overscale alarm condition for a period
of about 12 seconds�
• Operating the XNX EC sensor at extended temperature ranges for a prolonged time period exceeding 12 hours my cause deterioration in the sensor performance and shorten
sensor life� Extended temperature range for XNX EC sensors is -40°C ot -20°C�
• Response times may increase at lower temperatures�
• FM performance verification is limited to the requirements of the standards identified in Table 6�3 for each cartridge�
• Contact Honeywell Analytics for additional data or details�
Section 6 - Specifications
Range
Lower
Default
Incre-
Alarm
Range
ments
Limit
Full Scale)
n/a
23.0% Vol
n/a
5.0%Vol
15.0 ppm
0.1 ppm
5.0 ppm
300 ppm
100 ppm
30 ppm
sensor: The output from the O
2
sensor changes <±0�4%Vol� However, the oxygen sensor shows transient behavior when subjected to a rapid change in
2
Lower
Lower
Selectable
Explosive
Zero
Default
Detection
Cal Gas
Limit
Deviation
Cal Point
Limit
Range
(% Vol)
20.9 %Vol
5% Vol
n/a
n/a
20.9 %Vol
(fixed)
30 to 70%
1.5 ppm
n/a
-2.5 ppm
10 ppm
of the
selected full
scale range
15 ppm
na/
-25 ppm
100 ppm
sensor has pressure effects of <0�1% change of output per % change in pressure� When the barometric
2
131
Operating Temperature
Response
Response
Time (T50)
Time (T90)
Accuracy
sec
sec
Min
T20 <10
<30
<+/-0.5 %Vol
-30°C / -34°F 55°C /131°F
2 ppm or 10% of
<20
<30
reading, whichever is
-40°C / -40°F 55°C / 131°F
greater
<15
<30
See footnote 1
-40°C / -40°F 55°C / 131°F
Operating Humidity
Max
Min
Max
15% RH
90% RH
15% RH
90% RH
15% RH
90% RH