Honeywell XNX Technical Manual page 119

Universal transmitter
Hide thumbs Also See for XNX:
Table of Contents

Advertisement

XNX Universal Transmitter
Appli-
Fault
Description
cable
Sensors
Internal 4-20 mA
monitoring circuit
F149
All
communication
failure
mA Output Monitor
F150
Communications
All
Watchdog Error
Sensor Module
F151
ECC
Type Changed
Option Module
F152
All
Configuration Error
Signal/Data mis-
Optima,
F153
match error on IR
Excel
personality
mA Input Diagnos-
Optima,
F154
tic Failure
Excel
Generic mA Sensor
Generic
F155
Type Error
mA
mV Current Control
F156
mV
Failure
F157
Sensor Drift Fault
ECC, mV
Sensor/Personal-
F158
ity Part Number
All
mismatch
Section 5 - Warnings/Faults
Latching / Non-
Frequency of
Latching
Diagnostic
Non-latching
3�366 seconds
Non-latching
138 us
Non-latching
2 seconds
Only at powerup
or every 125 ms
Latching
when no option
board detected
Non-latching
2 seconds
5 minutes after
power up and
Latching
then every 8
hours
Non-latching
2 seconds
Non-latching
Main loop x16
Non-latching
2 seconds
"ECC & mV: main
loop x2;
Non-latching
Optima & Excel: 2
seconds"
Event History Data
0
Contact Honeywell Analytics' Service Department�
Communication error
Contact Honeywell Analytics' Service Department�
count
Module type: 0=None,
For ECC: Perform Accept New Sensor function, if problem persists
1=ECC, 2=mV,
contact Honeywell Analytics' Service Department� For others, contact
3=Excel, 4=Optima,
Honeywell Analytics' Service Department�
5=Generic mA
Option module ID:
0=None, 1=Founda-
Confirm option properly installed, reconfigure unit�
tion Fieldbus, 2=Mod-
bus, 3=Relay
Check wiring to Optima/Excel� In particular, check the white wire be-
Digital sensor reading
tween XnX and Optima/Excel� Note: power must be cycled to reset F153
after correcting the cause�
Input mA
Contact Honeywell Analytics' Service Department�
Indicates that mA input from sensor is less than 3 mA� Check wiring
between XNX and sensor� Also check the switches S3 and S4 are set
Input mA
correctly� If the switch settings need to be change, power down the XNX
before changing the switch settings� If wiring and switches are okay,
replace sensor�
constant current A/D
Check that correct mV sensor type is selected� Check wiring between
input mV
XNX and sensor� If sensor type and wiring are okay, replace sensor�
Current baseline
Perform zero calibration� If problem persists, replace sensor�
Entire personality
Check that installed option matches the option indicated in the XNX part
part #
number, check wiring to Optima/Excel�
117
Action For Resolution