5.1 Warning Messages
Applicable
Warning
Description
Sensors
XNX
24 VDC
®
W001
All
Supply Bad
XNX Tempera-
W002
All
ture Error
Simulated
W003
All
Warning/Fault
Sensor Tem-
Optima
perature Error
W005
Sensor Tem-
Excel
perature Error
Sensor Tem-
ECC
perature Error
Negative Drift
ECC, mV
W006
Optima,
Negative Drift
Excel
Calibration
W007
All
Required
Section 5 - Warnings/Faults
Frequency of Diag-
Latching / Non-Latching
nostic
Non-latching
2 seconds
Non-latching
2 seconds
Non-latching
Enabled by user
XNX polls sensor
every 2 seconds,
Non-latching
diagnostic fre-
quency controlled
by sensor
XNX polls sensor
every 2 seconds,
Non-latching
diagnostic fre-
quency controlled
by sensor
Non-latching
2 seconds
Non-latching
2 seconds
XNX polls sensor
every 2 seconds,
Non-latching
diagnostic fre-
quency controlled
by sensor
Non-latching
2 seconds
Event History Data
XNX supply
Check wire of 24V power supply to XNX as well as power supply opera-
voltage x1000
tion�
Check location for heat sources� Fit with sunshade or other protection�
XNX tempera-
Change location of XNX� Check temperature in Info->Transmitter Status
ture (Celsius)
to ensure temperature is being measured properly�
0
Performing an alarm/fault reset will clear all simulation�
Sensor fault or
Check location for heat sources� Fit with sunshade or other protection�
warning code
Change location of XNX� Check temperature in Info->Sensor Status to
(Note 4)
ensure temperature is being measured properly�
Sensor fault or
Check location for heat sources� Fit with sunshade or other protection�
warning code
Change location of XNX� Check temperature in Info->Sensor Status to
(Note 4)
ensure temperature is being measured properly�
Check location for heat sources� Fit with sunshade or other protection�
Sensor temper-
Change location of XNX� Check temperature in Info->Sensor Status to
ature (Celsius)
ensure temperature is being measured properly�
Raw gas con-
Check sensor location for external interference� Perform zero calibra-
centration of
tion� If problem persists after zero calibration and no interference exists,
sensor
replace sensor�
Check sensor location for external interference� Perform zero calibra-
Sensor fault or
tion� If problem persists after zero calibration and no interference exists,
warning code
replace sensor�
Number of days
remaining until
Time since the last span calibration has exceeded a defined limit� Per-
calibration ex-
forming a successful span calibration will clear the condition� The limit is
pires, negative
the user-defined calibration interval� W007 can be disabled by setting the
= number of
calibration interval to 0�
days expired
108
Action For Resolution