Test Chromatogram - Agilent Technologies 1100 Series Reference Manual

Variable wavelength detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Suggested Actions

Test Chromatogram

A pre-defined test chromatogram on the VWM board is processed through ADC like
normal signals from the diodes and can be used to check the ADC and the data handling
system. The signal is available at the analog output and on the GPIB.
The run time of the chromatogram is depending on the setting for response time (peak
NOTE
width). If no stop time is set the chromatogram will repeat continuously.
Table 6
Response time
0.06 sec
0.12 sec
0.25 sec
0.50 sec
1.00 sec
2.00 sec
4.00 sec
8.00 sec
Control Module
1 Select the response time and stop time according to
1100 Series Variable Wavelength Detector Reference Manual 71
Exchange the filter motor
Exchange the filter position sensor.
Exchange the VWM board.
Test Chromatogram Settings
Troubleshooting and Test Functions
Set Run Time
0.8 min
0.8 min
0.8 min
0.8 min
1.6 min
3.2 min
6.4 min
12.8 min
Table
6.
3