Honeywell Satellite XT Manual To Operations page 24

R version
Table of Contents

Advertisement

Mantenimiento - Sección 4
Sensor Service
To perform a sensor replacement with a new sensor,
which must have the same part number; see section
8 for Sensor Information. Use the <set> key to enter
and follow the dialogue displayed at the bottom line
of the screen. When the instrument has finished the
sensor service procedure, it will automatically enter
the Monitoring Mode.
Remove Sensor
Remove the sensor presently installed and actuate the
<set> key to proceed.
Replacement
Insert the new sensor and actuate the <set> key to
continue.
Load New Data ?
If you want the data of the new sensor to be loaded,
confirm this by actuating the <set> key.
Loading Data !
The new data are now loaded from the sensor into the
instrument's internal memory.
<Sensor Warm-Up Condition>
A sensor warm-up is performed and this screen is
shown until the displayed value is zero.
This does not apply for oxygen sensors.
The warm-up time required depends on the type of
sensor.
<Normal Monitoring Condition>
The instrument will automatically switch to the
Monitoring Mode when the sensor service procedure
is finished.
Satellite XT R Version
4-3