Invacare® Platinum™ Mobile

8 Technical Data

8.1 Technical Description

The Invacare Platinum Mobile uses a molecular sieve and pressure
swing adsorption methodology to produce the oxygen gas output.
Ambient air enters the device, is filtered and then compressed. This
compressed air is then directed into a nitrogen adsorbing sieve bed.
Concentrated oxygen exits the opposite end of the sieve bed and is
directed into an oxygen reservoir from which it is delivered to the
patient.
The oxygen purity level of the output gas ranges from 87% to 95.6%.
The oxygen is delivered to the patient through the use of a nasal
cannula. A pulse dose delivery method is used. The concentrator
detects the start of patient inhalation and delivers a measured pulse
of oxygen. No further oxygen is delivered until the next patient
inhalation is detected. The volume of oxygen delivered each minute is
a fixed amount based on the selected pulse flow setting. The volume
of each oxygen pulse will vary with the patient's breath rate such that
the fixed minute volume is maintained.
The Invacare Platinum Mobile is capable of operation by the patient in
a home, institution, vehicle, or other environments outside the home.
Device standard power options include an AC to DC switching
power adapter operating from an AC power outlet (100–240VAC,
50–60 Hertz nominal), a DC power cable operating from an
accessory outlet typically found in a vehicle type environment (12
VDC nominal), and up to two rechargeable batteries.
No specific product knowledge or training is required to operate the
product other than what is contained in this manual.
Service information will be available on request to qualified technical
personnel ONLY.
60

8.2 Specifications

Specifications for Optimal Performance
Operating Temperature:
Relative Humidity:
Transport/Storage Temperature:
Electrical:
Altitude:
Tubing and Cannula:
Environment:
Time of Operation:
41°F to 95°F (5°C to 35°C)
15% to 60%
-13°F to 122°F (-25°C to
50°C)
Allow unit to warm, or cool, to
operating temperature range
before using.
Do not use an extension cord.
Up to 10,000 ft (3048 m)
above sea level.
4 to 7 ft (1.2 to 2.1 m) cannula
with crush resistant tubing
(DO NOT pinch).
Smoke, pollutant and fume
free. No confined spaces
(example: no closets).
Up to 24 hours per day when
connected to an external
power source.
1194969-A-00