5.6 Cleaning and Disinfection Between Patients

Cleaning and disinfection must be performed between patients and as
necessary.
WARNING!
Risk of Injury or Damage
To prevent injury from infection or damage to
concentrator:
– Only qualified personnel should perform cleaning
and disinfection of the oxygen concentrator and
accessories between patients.
Follow these instructions to eliminate possible pathogen
exchange between patients due to contamination of
components or accessories.
1.
Dispose of and replace all patient side accessories not suitable
for multiple patient use, including but not limited to: cannula
and oxygen tubing.
2.
Clean concentrator and accessories as described in Maintenance.
1194969-A-00
3.
Disinfect the surfaces of the concentrator and accessories using
Clorox® 4 in One Disinfectant & Sanitizer or substantially
equivalent product.
WARNING!
Risk of Injury or Damage
– Follow the cleaning/disinfectant product
manufacturer's instructions for use, handling,
storage and disposal of the product.
– Use of inappropriate disinfecting products may
degrade device materials and negatively affect
device safety and performance.
If the accessories, such as the carry bag, are unable to
be cleaned or disinfected, discard the accessory and
replace.
4.
Perform the tasks on Preventive Maintenance Checklist.
5.
Before repackaging and distribution to new patient, ensure
packaging contents contain the concentrator, battery, carry bag
with shoulder straps, AC power adapter, AC power cord, DC
power cable, labels and user manual.