Field Calibration

The field calibration on this unit is done using the Field Calibration System.
Periodic field calibration is recommended to maintain measurement precision.
N
ote
• Always carry out field calibration when changing lenses.
• The Field Calibration System that corresponds to the type of lens the unit is equipped with must be used when con-
ducting field calibration.
• This unit and the Field Calibration System are paired, so it is not possible to correctly calibrate it when the serial
numbers differ. Use the unit and Field Calibration System combination that have the same serial number.
M
emo
The type of lens the unit is equipped with and the serial number of the unit can be confirmed in the Status information.
R
For details on displaying the status information, refer to page 41.
ef.

Field Calibration System

The Field Calibration System is comprised of the Unit Section which contains the Field Calibration System,
which corresponds to the type of lens on the unit, and the Frame Section, which maintains a facing position at
a set distance from the unit.
• Parts Nomenclature
Lock Lever B
Lock Lever A
32
Handle B
Support
Section
Rubber Foot (×6): With Height Adjustment Mechanism
The rubber feet adjustment mechanism has a structure
*
shown in the drawing below.
[Specifications]
• Lowest height: 43mm
• Adjustable range: Max. 10mm (with no preventive structure)
Handle A
Unit Section (Field Calibration System)
Equipped with field calibration systems for the
*
WIDE lens type and MIDDLE/TELE lens type.
Unit Section Cover
Unit Attachment Section
Frame Section
(Frame for Calibration System)
(Frame section)
Stopper
Lock nut
Rubber foot block
Main Unit Attachment
Section

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid 9i

Table of Contents