Konica Minolta VIVID 9i/VI Instruction Manual page 41

Non-contact 3d digitizer
Table of Contents

Advertisement

6
Check that field calibration was successful.
If you are using field calibration results to measure the dimensions of the field calibration system, the
software shows the differences from the measurements made at time of shipping. Differences are shown
for four places: X1, Y1, X2, and Y2.
• If field calibration was not successful, one of the following messages appears. Proceed in accordance
with the displayed message.
Message
Field Calibration:
Scan Error
Field Calibration:
Evaluate Error
Field Calibration:
Calibrate Error
7
Click the [Cancel] button.
The dialog closes, and field calibration is completed.
Meaning
Failed to recognize
• Remove the laser barrier and the receiving lens cap.
the field calibration
• Check that you are using the correct system type for the lens
system.
used.
• Check that foreign matter is not in the "Calibration Area".
• Check that there is no effect of light from outside.
• Be sure that the Field Calibration Chart unit is correctly
mounted to the Frame section.
• Check that the image of the Field Calibration Chart is cor-
rectly displayed in the view angle of the finger of the VIVID
9i/VI-9i or the Work window of the dialog box of the Poly-
gon Editing Tool.
Failed to correctly
• Check that you are using the correct system type for the lens
measure
the
field
used.
calibration system.
• Check that foreign matter is not in the "Calibration Area".
• Check that there is no effect of light from outside.
• Be sure that the Field Calibration Chart unit is correctly
mounted to the Frame section.
• Check that the image of the Field Calibration Chart is cor-
rectly displayed in the view angle of the finger of the VIVID
9i/VI-9i or the Work window of the dialog box of the Poly-
gon Editing Tool.
Adequate field cali-
• Remove the lens, and then set it again correctly.
bration results can-
• Check that foreign matter is not in the "Calibration Area".
not be obtained.
• Check that there is no effect of light from outside.
• Be sure that the Field Calibration Chart unit is correctly
mounted to the Frame section.
• Check that the image of the Field Calibration Chart is cor-
rectly displayed in the view angle of the finger of the VIVID
9i/VI-9i or the Work window of the dialog box of the Poly-
gon Editing Tool.
• When multiple units of the instrument are used, check to
make sure that the Field Calibration System matches the
serial number of the instrument.
• If there is any dirt on the laser window or the receiving lens,
clean these as explained on page 3.
• Clean off any dust on the field calibration system.
• If the field calibration system is scratched or dented, contact
a KONICA MINOLTA SENSING authorized service facil-
ity.
• If the same error recurs when you retry the field calibration,
contact a KONICA MINOLTA SENSING authorized ser-
vice facility.
Field Calibration
What To Do
39