General

Effective mAs

In sequential scanning, the dose (D
patient is the product of the tube current-time (mAs)
and the CTDI
w
D
= D
x mAs
seq
CTDI w
In spiral scanning, however, the applied dose (D
is influenced by the "classical" mAs (mA x Rot Time) and
in addition by the Pitch Factor. For example, if a Multi-
slice CT scanner is used, the actual dose applied to the
patient in spiral scanning will be decreased when the
Pitch Factor is larger than 1, and increased when the
Pitch Factor is smaller than 1. Therefore, the dose in
spiral scanning has to be corrected by the Pitch Factor:
D
= (D
x mA x Rot Time)/Pitch Factor
spiral
CTDI w
To make it easier for the users, the concept of the
"effective" mAs was introduced with the SOMATOM
Multislice scanners.
The effective mAs takes into account the influence of
pitch on both the image quality and dose:
Effective mAs = mAs/Pitch Factor
To calculate the dose on the SOMATOM Emotion 6,
you simply have to multiply the CTDI
the effective mAs of the scan:
D
= D
x effective mAs
spiral
CTDI w
For spiral scan protocols, the indicated mAs is the effec-
tive mAs per image. The correlation between tube
current mA and effective mAs of spiral scans on a Multi-
slice CT scanner is given by the following formula:
Effective mAs = mA x RotTime/Pitch Factor
Pitch Factor =
effective mAs
mA =
RotTime
where collimated Slice refers to the collimation of one
detector row, and nrow is the number of used detector
rows.
14
per mAs:
Feed/Rot
nrow x Slice collimation
x Pitch Factor
) applied to the
seq
spiral
per mAs with
W
)