Abdcta06S - Siemens SOMATOM Emotion 6 Application Manual

Table of Contents

Advertisement

Children

AbdCTA06s

Indications:
For abdominal CT Angio studies, e. g. vascular abnor-
malities, aneurysms, etc., using a 0.6 sec. rotation time.
A range of 15 cm will be covered in 11.2 sec.
kV
Effective mAs
Slice collimation
Slice width
Feed/Rotation
Rotation time
Kernel
Increment
Image order
* The mAs should be adjusted to the body weight
of the child.
Body weight
< 15 kg
15 – 24 kg
25 – 34 kg
35 – 44 kg
45 – 54 kg
For the 2
nd
reconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a sagittal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link
them to the series.
66
AbdCTA
80
*
1.0 mm
3.0 mm
9.0 mm
0.6 sec.
B41s
3.0 mm
cr-ca
kV
mAs
80
40
80
60
80
80
80
100
80
130
2
nd
reconstr.
1.25 mm
B31s
0.8 mm
CTDI
Vol
(mGy)
1.4
2.1
2.8
3.5
4.6

Hide quick links: