Spineroutineseq - Siemens SOMATOM Emotion 6 Application Manual

Table of Contents

Advertisement

Spine

SpineRoutineSeq

Indications:
Sequence mode for spine studies, e. g. prolapse,
degenerative changes, trauma, tumors etc.
This protocol contains three ranges:
L3-L4, L4-L5, L5-S1.
The scan length of each range is 2.8 cm
Three different typical gantry tilts are predefined:
for L3-L4: 0°, for L4-L5: +5° and for L5-S1: +15°
kV
mAs
Slice collimation
Slice width
Feed/Scan
Rotation time
Kernel
Cycle time
Image order
CTDI
Vol
176
SpineRoutineSeq
L3-4
L4-5
130
280
280
1.0 mm
2.0 mm
6.0 mm
1.5 sec.
B31s
3.0 sec.
cr-ca
36.9 mGy 36.9 mGy 42.2 mGy
L5-S1
320

Hide quick links: