Siemens SOMATOM Emotion 6 Application Manual page 163

Table of Contents

Advertisement

Abdomen
kV
Effective mAs
Slice collimation
Slice width
Feed/Rotation
Rotation time
Kernel
Increment
Image order
CTDI
Vol
nd
For the 2
reconstruction the Autoload into the
MIPthin Range on the 3D Card is activated. The images
will be automatically loaded into 3D, MIPthin, and
sagittal MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
MIPthin: AngioRoutine
Image thickness
Distance between images
Number of images
Contrast medium IV injection
Start delay
Flow rate
Total amount
* CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.
Do not administer oral contrast medium, as this
impairs the editing of MIP/SSD/VRT images. Use water
instead if necessary.
AngioRoutine
110
90
1.0 mm
5.0 mm
9.0 mm
0.6 sec.
B31s
5.0 mm
cr-ca
8.2 mGy
20 – 25 sec.*
3.0 – 3.5 ml/sec.
100 – 120 ml
2
nd
reconstr.
1.25 mm
B31s
0.8 mm
sagittal
3
3
60
163