Neck
kV
Effective mAs
Slice collimation
Slice width
Feed/Rotation
Rotation time
Kernel
Increment
Image order
CTDI
Vol
nd
For the 2
reconstruction the Autoload into the MPR
Range on the 3D Card is activated. The images will
be automatically loaded into 3D, MPR, and a sagittal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
MPRthick: NeckThinSlice
Image thickness
Distance between images
Number of images
Contrast medium IV injection
Start delay
Flow rate
Total amount
This protocol can also be used in case of suspected
tumor of the hypopharynx and larynx. It is recommen-
ded to acquire the images in E-phonation, and recon-
struct the images with 1 mm slice width and increment
of 0.7 mm for performing MPRs. This would help
differentiate tumor lesions from normal pharyngeal
mucosa.
NeckThinSlice
130
130
1.0 mm
5.0 mm
6.0 mm
0.8 sec.
B50s
5.0 mm
cr-ca
17.2 mGy
45 sec.
3 ml/sec.
100 ml
2
nd
reconstr.
1.25 mm
B50s
0.8 mm
sagittal
3
3
20
101