Siemens SOMATOM Emotion 6 Application Manual page 115

Table of Contents

Advertisement

Thorax
kV
Effective mAs
Slice collimation 1.0 mm
Slice width
Feed/Rotation
Rotation time
Kernel
Increment
Image order
CTDI
Vol
nd
For the 2
reconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
MIPthin: Embolism
Image thickness
Distance between images
Number of images
Contrast medium IV injection
Start delay
Flow rate
Total amount
* CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the pulmonary
trunc with triggering threshold of 120 HU, or use
manual triggering.
Embolism
130
70
3.0 mm
10.8 mm
0.8 sec.
B41s
3.0 mm
cr-ca
9.2 mGy
12 – 15 sec.*
4 ml/sec.
80 – 100 ml
2
nd
reconstr.
1.25 mm
B41s
0.8 mm
coronal
3
3
35
115