Siemens SOMATOM Emotion 6 Application Manual page 187

Table of Contents

Advertisement

Extremities
For the 1
st
reconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset, adapt the parameters to your needs and link them
to the series.
MIPthin: AngioRunOff
Image thickness
Distance between images
Number of images
Contrast medium IV injection
Start delay
Flow rate
Total amount 120 – 150 ml
* CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.
If Topo length 1024 mm is not long enough, you can
also choose the Topo length of 1500 mm.
Position the patient as feet first. Bend the feet together
if necessary.
25 – 30 sec.*
3.0 – 3.5 ml/sec.
coronal
3
3
35
187