Error Messages
The following is a list of the 5080A Calibrator error messages. The error message format
is shown in Table C-1.
Error Number
0 to 65535
Table C-1. Error Messages and Format
(Message Class : Description)
QYE Query Error, caused
by a full input buffer,
unterminated action or
interrupted action
DDE Device-Specific Error,
caused by the 5080A due to
some condition, for
example, overrange
EXE Execution Error,
caused by an element
outside of, or inconsistent
with, the 5080A capabilities
CME Command Error,
caused by incorrect
command syntax,
unrecognized header, or
parameter of the wrong type
0
(QYE:)
1
(DDE:FR)
100
(DDE:FR D)
102
(DDE:FR D)
104
(DDE:FR D)
Error Messages
F Error is displayed on the
front panel as it occurs
R Error is queued to the
remote interface as it occurs
S Error causes instrument to
go to Standby
D Error causes instrument
returns to the power up
state
(none) Error is returned to
the initiator only (i.e., local
initiator or remote initiator)
No Error
Error queue overflow
Inguard not responding (send)
Lost sync with inguard
Hardware relay trip occurred
Appendix C
Text Characters
Up to 36 text
characters
C-1