Error Messages

The Product display shows error messages when the
Product fails to meet certain conditions or there is a
Error Message
Too many errors
Cannot read file
Cannot recall procedure
Cannot store results
Table 8. Error Messages
Explanation
Too many errors, the error queue overflow.
Failed to read file from file system.
Failed to recall downloaded procedure.
Failed to store task results.
Automatic Pressure Calibrator
problem with the configuration of the Product or test. See
Table 8 to troubleshoot these issue.
Potential Root Cause and Solution
Clear the errors with the Product user
interface
1. File corrupted, delete the file and try
again.
2. Data flash memory failure, the main
PCA needs service.
1. The downloaded procedure contains an
illegal configuration.
2. File corrupted, download the procedure
again.
1. Not enough space, delete some files
and run the task again.
2. Delete all results and run the task again.
Maintenance
67