Chapter 4 Remote Interface Reference
Calibration Commands
Calibration Commands
For an overview of the calibration features of the function generator,
refer to "Calibration Overview" in chapter 3 starting on page 123.
For a detailed discussion of the function generator's calibration
procedures, refer to chapter 4 in the Agilent 33250A Service Guide.
CALibration:SECure:STATe {OFF|ON},<code>
CALibration:SECure:STATe?
Unsecure or secure the instrument for calibration. The calibration code
may contain up to 12 characters. The :STAT? query returns "0" (OFF) or
"1" (ON).
CALibration:SETup <0|1|2|3| . . . |115>
CALibration:SETup?
4
Configure the function generator's internal state for each of the
calibration steps to be performed. The :SET? query reads the calibration
setup number and returns a value from "0" to "115".
CALibration:VALue <value>
CALibration:VALue?
Specify the value of the known calibration signal as outlined in the
calibration procedures in the Agilent 33250A Service Guide. Use the
CAL:SET command to configure the function generator's internal state
for each of the calibration steps to be performed. The :VAL? query
returns a number in the form "+1.0000000000000E+01".
CALibration?
Perform a calibration of the instrument using the specified calibration
value (CAL:VAL command). Before you can calibrate the function
generator, you must unsecure it by entering the correct security code.
Returns "0" (PASS) or "1" (FAIL).