Getting Started
Programming
Example
# This example requires a 2 channel 81150A or 2 channel 81160A to run
without errors.
# Reset the instrument to start from a defined, default status.
*RST
# Switch off the automatic display update to increase programming
# speed
:DISPlay OFF
# Enable channel coupling.
# Initial coupling is done from channel 1 to channel 2
:TRACk:CHANnel1 ON
# Set the output function to pulse on channel 1
# Due to the channel coupling, this will also be applied to channel 2
:FUNCtion1 PULSe
# Setup Channel 1
# - Frequency 1MHz (will also be applied to channel 2)
# - Delay 0s
# - Pulse width 500 ns
# - Lead edge 75% of pulse width
# - Trailing edge 50% of pulse width
# - Offset 0V
# - Amplitude 3Vpp
:FREQuency1 1MHZ
:PULSe:DELay1 0S
:PULSe:WIDTh1 500NS
:PULSe:TRANsition:HOLD WRATio
:PULSe:TRANsition1 75PCT
:PULSe:TRANsition1:TRAiling 50PCT
:VOLT1:OFFSet 0V
:VOLT1 3VPP