Signal
Source
I/O Connector
© National Instruments Corporation
which has an internal 2 MHz clock. Refer to the DAQCard-1200
Register-Level Programmer Manual for programming information.
The GATE, CLK, and OUT signals for counters B1 and B2 are available at
the I/O front connector. The GATE and CLK pins are internally pulled up
to +5 V through a 100 k resistor. Refer to Appendix A, Specifications, for
signal voltage and current specifications.
You perform pulse and square wave generation by programming a counter
to generate a timing signal at its OUT output pin. Perform event counting
by programming a counter to count rising or falling edges applied to any of
the 82C53 CLK inputs. You can then read the counter value to determine
the number of edges that have occurred. You can gate counter operations on
and off during event counting. Figure 3-17 shows connections for a typical
event-counting operation in which a switch is used to gate the counter on
and off.
Switch
13

Figure 3-17. Event-Counting Application with External Switch Gating

+5 V
10 k
CLK
GATE
Counter (from Group B)
DGND
DAQCard-1200
3-27
Chapter 3
Signal Connections
OUT