Chapter 4
Theory of Operation
Note
If you use external conversions for single-channel acquisitions, the maximum
sampling rate is 90 kS/s.
DAQCard-1200 User Manual
For multiple-channel data acquisition, observing the data acquisition rates
in Table 4-2 ensures 12-bit accuracy.

Table 4-1. Analog Input Settling Time Versus Gain

Gain Setting
1, 2, 5, 10
20
50
100

Table 4-2. DAQCard-1200 Maximum Recommended Data Acquisition Rates

Acquisition Mode
Single channel
Multiple channel
The recommended data acquisition rates in Table 4-2 assume that voltage
levels on all the channels included in the scan sequence are within range for
the given gain and are driven by low-impedance sources.
Settling Time Recommended
10 s max
12 s typ, 15 s max
25 s typ, 30 s max
60 s typ, 80 s max
Gain Setting
1, 2, 5, 10, 20, 50, 100
1, 2, 5, 10
20
50
100
4-10
Sampling
Rate
100.0 kS/s
100 kS/s
83.3 kS/s
40.0 kS/s
16.6 kS/s
© National Instruments Corporation