Single-Ended Connections For Grounded Signal Sources (Nrse Configuration); Figure 3-5. Single-Ended Input Connections For Floating Signal Sources - National Instruments DAQCard-1200 User Manual

Multifunction i/o card for the pcmcia bus
Hide thumbs Also See for DAQCard-1200:
Table of Contents

Advertisement

Chapter 3
Signal Connections
+
Floating
Signal
V s
Source
Front I/O Connector
DAQCard-1200 User Manual
ACH 0
1
ACH 1
2
3
ACH 2
ACH 7
8
AISENSE/AIGND
9
AGND
11
DAQCard-1200 in RSE Configuration

Figure 3-5. Single-Ended Input Connections for Floating Signal Sources

Single-Ended Connections for Grounded Signal Sources
(NRSE Configuration)
If you measure a grounded signal source with a single-ended configuration,
you must configure the DAQCard-1200 in the NRSE input configuration.
The signal is connected to the positive input of the DAQCard-1200
instrumentation amplifier and the signal local ground reference is
connected to the negative input of the DAQCard-1200 instrumentation
amplifier. Therefore, you must connect the ground point of the signal to the
AISENSE pin. Any potential difference between the DAQCard-1200
ground and the signal ground appears as a common-mode signal at both the
positive and negative inputs of the instrumentation amplifier and is
therefore rejected by the amplifier. On the other hand, if the input circuitry
of the DAQCard-1200 is referenced to ground, such as in the RSE
configuration, this difference in ground potentials appears as an error in the
measured voltage.
Figure 3-6 shows how to connect a grounded signal source to a
DAQCard-1200 configured in the NRSE configuration. For configuration
instructions, see the
Analog Input Configuration
Installation and
Configuration.
+
3-12
+
Measured
V m
Voltage
section in Chapter 2,
© National Instruments Corporation