SECTION 2—COMPONENT REPLACEMENT
SECTION 2—COMPONENT
REPLACEMENT
To insure proper operation and avoid voiding unit warranties only use Invacare supplied 
components.

Cleaning the Gross Particle Filter

NOTE: For this procedure, refer to FIGURE 2.1.
NOTE: The gross particle filter should be cleaned or replaced according to the Transportable 
Oxygen Concentrator Preventive Maintenance Record on page 23.
DO NOT operate the Transportable Oxygen Concentrator without filters installed.
DO NOT clean the filter with alcohol and alcohol based products (isopropyl alcohol), 
concentrated chlorine based products (ethylene chloride), or oil based products (Pine‐
Sol®, Lestoil®) or any other harsh chemical agents. Only use mild liquid dish detergent 
(such as Dawn™).
1. Squeeze thumb tabs on filter cover and 
remove from unit.
2. Lift out air intake filter.
3. Use a vacuum cleaner or wash filter 
with a mild liquid dish detergent (such 
as Dawn
) and water. Rinse 
thoroughly.
4. Thoroughly dry the filter and inspect 
for fraying, crumbling, tears and holes. 
A. Replace filter if damaged.
5. Reinstall the filter cover by placing tabs 
in slots, push to engage into place.
NOTE: Use only Invacare replacement part
number 1156863 (Gross Particle Filter) or
1156861 (Gross Particle Filter and Filter
Cover Kit) on the Transportable Oxygen
Concentrator.
SOLO
2
CAUTION
CAUTION
L
Thumb
Tabs
Filter
Cover
FIGURE 2.1 Cleaning the Gross Particle
10
Filter
Cover
Filter
Filter
Part No 1164898