SECTION 5—TROUBLESHOOTING
AND MAINTENANCE
Functional Test
NOTE: For this procedure, refer to FIGURE 5.1.
1. Install battery in unit.
2. Verify communication exists between the Transportable Oxygen Concentrator and the
battery (DETAIL "A") by momentarily pressing the "ON/OFF"
3. Connect Transportable Oxygen Concentrator to AC power source.
• Screen will light up displaying the AC power icon, battery %, bars on battery
symbol that correspond to the battery % and the charging icon if battery is not at
100% (DETAIL "B").
DETAIL "A"
DETAIL "B"
DETAIL "C"
4. Turn the Transportable Oxygen Concentrator on by pressing the "ON/OFF"
for approximately 2‐3 seconds, not to exceed 5 seconds.
NOTE: When the Transportable Oxygen Concentrator is turned on, the green LED will illuminate
to show unit is in operation.
5. Change setting to continuous flow. Refer to the Transportable Oxygen Concentrator
Owner's Manual for this procedure.
6. Allow the unit to run for a minimum of 8 minutes, then check the oxygen purity. Refer
to Checking O
Purity on page 13.