SECTION 5—TROUBLESHOOTING
AND MAINTENANCE

Functional Test

NOTE: For this procedure, refer to FIGURE 5.1.
1. Install battery in unit.
2. Verify communication exists between the Transportable Oxygen Concentrator and the 
battery (DETAIL "A") by momentarily pressing the "ON/OFF" 
3. Connect Transportable Oxygen Concentrator to AC power source.
• Screen will light up displaying the AC power icon, battery %, bars on battery 
symbol that correspond to the battery % and the charging icon if battery is not at 
100% (DETAIL "B"). 
DETAIL "A"
DETAIL "B"
DETAIL "C"
4. Turn the Transportable Oxygen Concentrator on by pressing the "ON/OFF" 
for approximately 2‐3 seconds, not to exceed 5 seconds.
NOTE: When the Transportable Oxygen Concentrator is turned on, the green LED will illuminate 
to show unit is in operation.
5. Change setting to continuous flow. Refer to the Transportable Oxygen Concentrator 
Owner's Manual for this procedure.
6. Allow the unit to run for a minimum of 8 minutes, then check the oxygen purity. Refer 
to Checking O
 Purity on page 13. 
2
Part No 1164898
SECTION 5—TROUBLESHOOTING AND MAINTENANCE
FIGURE 5.1 Functional Test