Replacing the Inlet Filter
NOTE: For this procedure, refer to FIGURE 2.2.
NOTE: The inlet filter should be replaced according to the Transportable Oxygen Concentrator
Preventive Maintenance Record on page 23.
1. Position Transportable Oxygen Concentrator flat on the backside of unit to locate the
inlet filter access door (DETAIL "A").
2. Using a #2 Phillips® screw driver, remove the retaining screw securing the inlet filter
access door to unit (DETAIL "B").
3. Remove inlet filter access door from unit.
4. Locate existing inlet filter and remove (DETAIL "C").
5. Install new replacement inlet filter.
6. Reinstall inlet filter access door by inserting tabs into the slots located on the unit.
7. Reinstall retaining screw to secure the inlet filter access door.
NOTE: Use only Invacare replacement Part Number 1157083 (Inlet Filter) on the Transportable
Oxygen Concentrator.
DETAIL "A"
Filter Access
Door
Part No 1164898
SECTION 2—COMPONENT REPLACEMENT
DETAIL "C"
FIGURE 2.2 Replacing the Inlet Filter
11
DETAIL "B"
Screw
Inlet Filter
™
SOLO
2