SECTION 4—ADMINISTRATIVE SETTINGS SCREEN
Toggling Auto Pulse
NOTE: For this procedure, refer to FIGURE 4.1 on page 14.
1. Enter administrative settings screen (DETAIL "B"). Refer to Accessing Administrative
Settings Screen on page 14 for this procedure.
2. Locate the Auto Pulse setting (DETAIL "B").
A. Press the Up/Increase
setting.
B. Press the Return/Highlight button to relocate highlight bar to the adjustable setting
on right side of screen.
3. Press the Up/Increase
4. Save changes by pressing the Return/Highlight
left side of screen.
NOTE: Changes are not saved until the highlight bar is relocated to the left side of the screen.
5. Press and hold the Return/Highlight
screen.
NOTE: The Transportable Oxygen Concentrator will automatically return to the main menu
operational screen if idle for a period of approximately 10‐15 seconds.
NOTE: Unit will remain in Auto Pulse mode until Auto Pulse is returned to "OFF".
™
SOLO
2
or Down/Decrease
or Down/Decrease
button to return to the main menu operational
16
button to highlight Auto Pulse
button to toggle between ON or OFF.
button to relocate highlight bar to
Part No 1164898